Saxsons Group — India's trusted nuclear medicine, radiotherapy, oncosurgery, dosimetry and cyclotron supplier since 1986

LUCY™ 3D QA Phantom — what the SRS end-to-end literature says.

Four real anchor citations across the workflow the Lucy 3D is bought for — the end-to-end (E2E) portion of an SRS / SRT / SBRT QA programme under the AAPM MPPG 9.a framework. Covers the direct peer-reviewed head-to-head (Sarkar 2016 · Lucy 3D vs a commercial SRS QA comparator), the practice guideline that requires an E2E phantom (Halvorsen 2017 · MPPG 9.a · 72-hour availability), the applied comparative E2E QA study quantifying film vs diode-array workflow trade-offs (Lim 2022) and the numerical commissioning-criteria thresholds a modern MLC-based SRS programme is measured against (Brown 2022). Every DOI + journal reference verified; all open-access URLs.

< 1 mmWinston-Lutz isocentre to hidden-target coincidence (Brown 2022)
≤ 72 hrAAPM MPPG 9.a E2E-phantom on-site availability requirement
γ > 90 %film γ (3 % / 1 mm) passing rate — SRS commissioning threshold
Filter
What the Sarkar 2016 head-to-head measured
comparator
Commercial comparator
modern commercial SRS QA phantom
vs
subject
Lucy 3D
Exradin A16 micro chamber + commercial SRS headframe + MR distortion inserts
Peer-reviewed comparison2016
Lucy 3D head-to-headDirect peer-reviewed SRS-phantom head-to-head — Exradin A16 micro chamber inside a commercial SRS headframe, with MR-signal-generator and grid-distortion inserts for MRI geometric QA. The only peer-reviewed comparison that names the Lucy 3D against a modern commercial SRS-QA comparator phantom

Head-to-head comparison of two commercial phantoms used for SRS QA

Journal of Radiosurgery + SBRT 4(3):213 – 223 (open access) — Sarkar, Huang, Huang, Rassiah-Szegedi, Salter

MPPG 9.a — what the SRS E2E phantom must do
Scope
appropriate to the SRS-SBRT scope of services
≤ 72 hr
on-site availability requirement
Same
staff continuity across every E2E step
MPPG 9.a is the modern practice guideline for medical-physics support of linac-based SRS / SRT / SBRT programmes. The E2E-phantom requirement is verbatim in the guideline; the Lucy 3D is a purpose-built way to meet it.
Practice guideline · MPPG 9.a2017
72 hrMPPG 9.a explicitly requires an E2E phantom appropriate to the SRS-SBRT scope of services on-site + available within 72 hours; each E2E step performed by the same staff. IROC-Houston is the only external-audit resource named. This is the framework paper the Lucy 3D is bought against

AAPM-RSS Medical Physics Practice Guideline 9.a — SRS-SBRT

Journal of Applied Clinical Medical Physics 18(5):10 – 21 (open access) — Halvorsen, Cirino, Das, Fairobent, Fontenla, Gopalakrishnan, Kabuka, Klein, Lo, Mahesh, Nash, Prisciandaro, Sharpe, Sherouse, Watkins (AAPM-RSS)

SRS E2E QA — film vs diode array — Lim 2022
γ 94 – 100 %
Anthropomorphic + radiochromic film
reference method; 6 h analysis; localisation within 0.5 ± 0.2 mm
vs
± 1.9 % vs TPS
Commercial SRS diode-array device
faster (2 h analysis); +0.7 mm localisation uncertainty vs film
Applied E2E · Film vs diode array2022
94 – 100 %Radiochromic-film γ (3 % / 1 mm) passing rate 94.0-100.0 % (reference method). Commercial diode-array E2E device within 1.9 % of the TPS-calculated dose but adds ~ 0.7 mm localisation-uncertainty vs film. Quantifies where the fast diode-array workflow is close-enough vs where the reference-film E2E cross-check still has to be done

Comparative study of SRS end-to-end QA processes of a diode-array device and an anthropomorphic phantom loaded with GafChromic XD film

Journal of Applied Clinical Medical Physics 23(9):e13747 (open access) — Lim, Kuo, Li, Li, Ballangrud, Lovelock, Chan (Memorial Sloan Kettering)

MLC SRS commissioning criteria — Brown 2022
< 1 mm
Winston-Lutz iso ↔ hidden-target coincidence
± 5 %
measured-chamber vs calculated dose
γ > 90 %
film γ (3 % / 1 mm) passing rate
These are the numerical acceptance criteria a modern MLC-based SRS programme is commissioned against. The Lucy 3D covers the Winston-Lutz + chamber-dose + film-γ evidence chain in one phantom.
MLC SRS commissioning criteria2022
< 1 mm · ± 5 % · γ > 90 %The three concrete commissioning-criteria thresholds an MLC SRS programme on a modern C-arm linac is measured against: coincidence of treatment isocentre and CBCT-aligned hidden target < 1 mm; measured-chamber to calculated dose within ± 5 %; film γ (3 % / 1 mm) > 90 %. Tighter than IROC 5 % / 3 mm — reflects the precision requirement of stereotactic delivery

Determination of commissioning criteria for multileaf-collimator, stereotactic radiosurgery treatments on Varian TrueBeam and Edge machines using a novel anthropomorphic phantom

Journal of Applied Clinical Medical Physics 23(6):e13581 (open access) — Brown, Fagerstrom, Beck, Holloway, Burton, Kaurin, Mahendra, Luckstead, Kielar, Kerns (NW Medical Physics Center)

Standards, Regulator + Manufacturer Reference Reading

AAPM SRS-SBRT + linac-QA task-group reports the phantom-based E2E workflow sits inside, plus manufacturer brochure + E2E whitepaper.

Literature Radar · auto-updated daily

Latest SRS end-to-end QA research

Fresh peer-reviewed papers on end-to-end stereotactic QA — phantom design + head-to-head studies, MPPG 9.a implementation, MLC SRS commissioning criteria, and multi-target single-isocentre / MR-guided workflows. Auto-updated from Europe PMC.

See all in radar →
Loading recent papers…