Saxsons Group

Medical Physicist's Dossier · myOSL™ Chip

Linearity, energy response, fading, angular dependence — the numbers, on one page.

A short, technical dossier for the medical physicist commissioning OSL in-vivo for the first time or switching from the discontinued Landauer nanoDot. The five panels below are internal-testing QA results from the manufacturer type-test programme — the kind of charts that decide whether a detector is worth taking to your linac. Peer-reviewed validation papers (Kowalski 2025, Davis 2025) live on the Knowledge Hub.

Data on this page is reproduced from the manufacturer internal type-test programme. Saxsons Group serves as the authorised Indian distributor and can arrange on-site evaluation against your own reference dosimeters.

TEST 01

Linearity — personal dose equivalent Hp(10)

Measured Hp(10) vs reference dose, 0 → 1 000 mSv.

Result

R² = 0.9994 · slope 0.9007

Linearity — personal dose equivalent Hp(10)

Interpretation

Across three orders of magnitude — from background-level (sub-mSv) to accident-tier (1 Sv) — the system's response is linear with R² = 0.9994. A near-unity slope (0.9007) means a single energy-independent calibration factor recovers the reference reading; no piecewise correction needed.

What it means for your workflow

Calibrate once, trust the curve across the full operational dose range.

TEST 02

Linearity — extended dose range (radiotherapy regime)

OSL signal (a.u.) vs delivered dose, 0 → 10 Sv (dosimeter #3520).

Result

r² = 0.9999 to 10 Sv

Linearity — extended dose range (radiotherapy regime)

Interpretation

The same linear behaviour holds an order of magnitude further up, into the radiotherapy in-vivo dose regime. r² = 0.9999 from zero to 10 Sv on a single dosimeter means BeO saturation and supralinearity aren't practical concerns at clinical doses.

What it means for your workflow

Safe for in-vivo entrance/exit checks at any clinical fractionation, including SBRT-level single fractions.

TEST 03

Energy response — tissue equivalence

(μ_en/ρ)_material / (μ_en/ρ)_tissue across 10 keV → 10 MeV photons. BeO compared against CaF₂, CaSO₄, Al₂O₃, LiF, Li₂B₄O₇ and air.

Result

Z_eff 7.21 · flat from ~50 keV upward

Energy response — tissue equivalence

Interpretation

BeO's effective atomic number (7.21) sits right next to soft tissue (7.35), so the mass-energy-absorption ratio stays near unity across the entire diagnostic and therapeutic photon range. Above ~50 keV the response is essentially flat — well within ±5 %. Al₂O₃:C (Z_eff ≈ 11) over-responds 2–3× in the kV diagnostic range; BeO does not.

What it means for your workflow

No energy-dependent correction needed for routine kV/MV photon dosimetry. Drop-in replacement for Al₂O₃:C nanoDots without a workflow redesign.

TEST 04

Signal fading — 15 min, 1 week, 3 months

Mean recovered dose at three read-out delays after a 0.35 mSv reference irradiation. Background subtracted.

Result

−0.16 % at 3 months

Signal fading — 15 min, 1 week, 3 months

Interpretation

After 15 minutes the deviation from reference is −0.37 %. After one week, −0.51 %. After three months, −0.16 % — i.e. there is no statistically meaningful fading on a routine quarterly read-out cadence. Read-out scheduling is therefore a logistics question, not a measurement-accuracy question.

What it means for your workflow

Quarterly badge cycles, mailed samples, and delayed-readout protocols all fit comfortably within the system's fade tolerance.

TEST 05

Angular response — front vs side irradiation

Polar response across 60°, 90°, 120° irradiation angles for ISO N-series narrow-spectrum beams (N60 → N200) plus Cs-137 and Co-60.

Result

< 20 % drop at 90° (side)

Angular response — front vs side irradiation

Interpretation

Side-on irradiation reduces the indicated value by less than 20 % relative to en-face — well inside the angular-dependence band specified in IEC personal-dosimetry test protocols. Across the full N60 → Co-60 energy range the spread is consistent, so a single angular correction handles real-world badge orientations.

What it means for your workflow

Passes IEC type-test angular requirements. Realistic body-mounted use does not need a per-angle correction.

Summary — at a glance

Metric Result Reference
Linearity (0 → 1 Sv) R² 0.9994 Hp(10) test, slope 0.9007
Linearity (0 → 10 Sv) r² 0.9999 extended single-dosimeter test
Fading (3 months) −0.16 % vs 0.35 mSv reference
Fading (1 week) −0.51 % σ = 1.35 %
Angular response (side vs front) < 20 % IEC type-test compliant
Tissue equivalence Z_eff 7.21 soft tissue 7.35
Detector material BeO 4.7 × 4.7 × 0.5 mm element
Housing 9.5 × 10 × 2 mm ABS light-tight

Independent peer-reviewed validation (Kowalski 2025 — Duke; Davis 2025 — UVA) confirms these characteristics against the discontinued Landauer nanoDot. Full citations on the Knowledge Hub.

Bring it to your linac

Audit our numbers against your own reference dosimeters.

Saxsons Group will arrange a trial set of myOSL Chip dosimeters and a reader so your physics team can replicate the linearity, fading and angular tests above in your own QA programme — before any commitment.