Medical Physicist's Notes · Ge-68 Line Source
PET normalization is not a once-a-year event. The full NEMA NU-2 normalization runs quarterly and after service; lighter daily / weekly checks catch detector drift before it contaminates the quarterly baseline. This post draws the four cadence bands and lists the drift modes that catch programmes between scheduled normalizations.
The four cadence bands
| Band | Trigger | Test | Pass | Fail action |
|---|---|---|---|---|
| Daily / morning | Before the first clinical PET acquisition of the day | Coincidence detector counts on the Ge-68 cylindrical phantom; sinogram uniformity check | Detector counts within ±5 % of yesterday's value; sinogram visually uniform | Hold clinical scanning. Repeat after detector warm-up; check for a tripped PMT module |
| Weekly | One scheduled slot per week (typical Monday morning) | Full uniformity + slice-sensitivity profile from a Ge-68 cylindrical or line source | Axial uniformity < 5 %, slice-sensitivity profile within tolerance | Investigate single-slice anomalies; consult service engineer if root cause is a detector / electronics fault |
| Quarterly | Once per quarter; aligned with cyclotron radiopharmacy QA | Full NEMA NU-2 normalization scan with rotating Ge-68 line source; per-crystal sensitivity map regenerated | New normalization file accepted; CT/PET co-registration verified; quantitative SUV calibration on F-18 cylinder within ±10 % | Quarantine the normalization until service engineer inspects detector electronics |
| After service | Detector block swap, gantry mechanical service, software upgrade, scanner relocation | Full NEMA NU-2 normalization + uniformity + sensitivity + SUV-cylinder cross-check | Pre-service / post-service comparison within stated tolerance; baseline reset | Hold clinical scanning until baseline is recovered and documented in the service log |
Source: AAPM TG-181 PET QA guidance; IAEA Human Health Series 27; NEMA NU-2-2018.
Drift modes
Source: AAPM TG-181; manufacturer-specific service guidance.