Saxsons Group

Per-patient selection · Breast Positioning

Eight selection factors — supine or prone per patient.

Per-patient supine vs prone breast-positioning selection is a clinical decision shaped by breast volume, dosimetric trade-offs, DIBH compatibility, patient tolerance and setup reproducibility. This page is the eight-factor selection matrix used at simulation to lock the per-patient workflow.

Factor Supine Prone
Breast volume Routine for ≤ B-cup; manageable up to C-cup with appropriate cushion Recommended for ≥ C-cup; mandatory consideration ≥ D-cup where supine breast crosses the midline
Ipsilateral lung dose V20 typically 12–18 % for tangential left-breast (Gy) V20 typically drops 30–50 % vs supine for larger breasts; lung kept out of the high-dose tangent path
Cardiac dose (left-side) Heart-mean-dose typically 2–5 Gy free-breathing; DIBH reduces to 1–2 Gy Heart-mean-dose 1–3 Gy free-breathing; less DIBH-dependent due to cardiac displacement
Skin dose Higher at the breast-axilla junction; skin-fold dose can drive moist desquamation Lower at the breast-axilla junction; breast tissue drapes away from the chest wall
Patient tolerance Most patients comfortable; lengthy DIBH protocols (≥ 20-min total breath-hold time) can be tiring More demanding posture; requires patient mobility + comfort assessment; some patients cannot tolerate prone for the fraction duration
Setup reproducibility Workflow standardised; daily CBCT verification well-established Requires breast-board angle indexing + cushion conformal hold; CBCT visualisation can be more challenging for some breast geometries
Combined with DIBH Standard combination; T-grip + breast cushion + indexed board hold the DIBH posture across fractions DIBH less commonly combined with prone; the prone posture displaces heart away from chest wall already
Bilateral or chest-wall Standard for chest-wall and bilateral breast workflows; tangential field arrangement is direct Not typically used for bilateral or chest-wall workflows; prone is single-breast geometry

How to use this matrix

At simulation, the radiation oncologist and physicist score each factor against the patient profile. Breast volume and ipsilateral lung dose typically drive the selection; cardiac dose for left-side cases shifts the balance toward DIBH-supine unless the prone advantage is large. Patient tolerance is the over-ride — a patient who cannot tolerate prone for the fraction duration is treated supine regardless of dosimetric advantage. The accessory family covers both modes, so the per-patient decision is purely clinical, not constrained by equipment.