Gamma and Fluorescence bi-modal detection
Features
- Automatic switching to most common radioisotopes (99mTc, 125I, 131I, 18F)
- Customized user setting of energy window
- Easy system set-up
- No need for calibration
- No drift of the probe in time
- 9 customized configurations + 1 default
Gamma module
With a single readout module and numerous gamma probe options, the Europrobe 3.2 is a unique system that fulfills all needs of per and post-operative detection as well as per-cutaneous localisation within 7 major clinical fields.
This solution allows sentinel node dual detection Gamma and Fluorescence.
Clinical applications
Breast cancer, Gynaecology, Head & Neck, Dermatology, Endocrinology, Nuclear medicine, Urology
Europrobe features
Main features
- Automatic switching to most common radioisotopes (99mTc, 125I, 131I, 18F)
- Customized user setting of energy window
- Easy system set-up
- No need for calibration
- No drift of the probe in time
- 9 customized configurations + 1 default
Technical specifications
- Maximum count: 30 000 cps
- Operating temperature range: +15˚C to +40˚C
- Height: 245 mm
- Width/Depth: 233 mm/270 mm
- Weight: 2 Kg
- Power consumption: Max 15 watts
- Operating power: 115/220 VAC F=50Hz
CsI probes
The multiple benefits of CSI:
- Works at ambient temperature
- Detects hight energies
- Works with a low bias voltage unlike the PMT solution (typically 20V)
- Available in wireless Bluetooth version
- Possibility of offering customized probes
Characteristics:
- Energy: 100 keV-1 MeV (Optimum efficiency: 140 keV-1 MeV)
- Isotopes: 99mTc (140 keV), 111In (171 & 245 keV) , 131I (364 keV), 18F (avec un collimateur, 511 keV)
- Medical grade stainless steel
CdTe probes
The multiple benefits of CdTe:
- Works at ambient temperature
- Detects low energies
- Offers excellent energy resolution at 140 keV ( 99mTc)
- Small probes for detection in small incisions with high accuracy
- Available in wireless Bluetooth version
- Possibility of offering customized probes
Characteristics:
- Energy: 20-300 keV (Optimum efficiency: 20-170 keV)
- Isotopes: 99mTc (140 keV),111In (171 & 245 keV ) ), 125I (27 keV)
- Medical grade stainless steel
Fluorescence module
An additional module to the Europrobe allows dual optical detection by fluorescence and radio-nucleotide for ganglion localization. The combined use of these two detection methods reaches almost a 100% detection rate. This dual fluorescence and radioisotope detection method within one device is a worldwild exclusive.
Main features
- Optional add-on module
- Allows dual Gamma and Fluorescence detection
- Selection of the type of detection through a foot-switch
- Simple and identical use to the Gamma procedure
- Ready to use at normal light conditions
- No specific installation required
- Improved discrimination and sensitivity
Technical specifications
- Suitable for of Indocyanine Green (ICG) detection
- Module height: 55mm
- Width / Depth: 233 mm / 270 mm
- Weight: 1 Kg
Dual probes
Dual detection probes: gamma and Fluorescence.
Opto-nuclear probes are designed to detect both isotopic markers
and fluorescent markers (ICG).
Characteristics :
- Energy: 20 – 300 keV (Optimum efficiency: 20 – 170 keV)
- Isotopes: 99mTc, 125I, 111In – Sensitivity of fluorescence detection: <1nmol of ICG
- Medical grade stainless steel